Articles | Volume 9, issue 2
https://doi.org/10.5194/acp-9-585-2009
https://doi.org/10.5194/acp-9-585-2009
27 Jan 2009
 | 27 Jan 2009

Ozone Monitoring Instrument spectral UV irradiance products: comparison with ground based measurements at an urban environment

S. Kazadzis, A. Bais, A. Arola, N. Krotkov, N. Kouremeti, and C. Meleti

Abstract. We have compared spectral ultraviolet overpass irradiances from the Ozone Monitoring Instruments (OMI) against ground-based Brewer measurements at Thessaloniki, Greece from September 2004 to December 2007. It is demonstrated that OMI overestimates UV irradiances by 30%, 17% and 13% for 305 nm, 324 nm, and 380 nm respectively and 20% for erythemally weighted irradiance. The bias between OMI and Brewer increases with increasing aerosol absorption optical thickness. We present methodologies that can be applied for correcting this bias based on experimental results derived from the comparison period and also theoretical approaches using radiative transfer model calculations. All correction approaches minimize the bias and the standard deviation of the ratio OMI versus Brewer ratio. According to the results, the best correction approach suggests that the OMI UV product has to be multiplied by a correction factor CA(λ) of the order of 0.8, 0.88 and 0.9 for 305 nm, 324 nm and 380 nm respectively. Limitations and possibilities for applying such methodologies in a global scale are also discussed.

Download
Altmetrics
Final-revised paper
Preprint